The Plasma Assay’s Sensitivity Depended on Its Validated Specimen Volume

The published plasma assay used a one-milliliter specimen in its preparation procedure. Its concentration sensitivity cannot automatically be assigned to a much smaller-volume version of the method.

Volume was part of the workflow

The protocol precipitated proteins from 1 mL plasma with 2 mL acetonitrile before extraction, evaporation and derivatization. Its reported lower quantification limits included 0.1 ng/mL for THC and 11-hydroxy-THC. Those values describe the complete tested procedure, not an instrument specification independent of specimen preparation.

Concentration and available amount differ

At the same concentration, a smaller specimen contains less analyte. Changing volume can also change recovery, matrix contribution and final extract composition. A miniaturized procedure may perform well, but that is an empirical validation question. The paper did not demonstrate that an arbitrary smaller aliquot retains all of its reported performance.

What a laboratory comparison requires

Review the volume actually processed, extraction and concentration steps, injection conditions and current lower-limit data. A different laboratory may have validated a lower-volume method; this publication neither certifies nor disproves it. The study concerns analytical quantification in plasma for research, with stated limits on forensic transfer. Sensitivity at a defined volume is not evidence of an individual’s dose or degree of impairment.

Related analysis: the plasma method’s intended research use.

Research cited

Andrenyak and colleagues (2017). Determination of THC, 11-hydroxy-THC, THC-COOH and Cannabidiol in Human Plasma using Gas Chromatography–Tandem Mass Spectrometry. DOI: 10.1093/jat/bkw136.

Spread the love


The National Black Lawyers

top 40 lawyers

civil trial law

Lawyers of Distinction

Loading...

Recent Blog Articles

Minor Breath Cannabinoids Were Not Detected in Every Early Sample

CBC and THCV showed promising early detection patterns, but neither was detected in every evaluated participant within the first hour. Their individual findings should not inherit the combined test’s reported 100% classification result. The marker counts were specific In the report’s table, CBC was detected in thirty-nine of forty participants…

Spread the love

Read Article: Minor Breath Cannabinoids Were Not Detected in Every Early Sample

Breath THC Half-Life Estimates Changed With the Sampling Interval

The reported breath THC half-life was much shorter immediately after smoking than at later sampling intervals. A two-point breath estimate was not a fixed whole-body elimination half-life. Different portions of the decline produced different values The report gave an average estimate of approximately 4.2 minutes over the zero-to-twenty-minute interval, compared…

Spread the love

Read Article: Breath THC Half-Life Estimates Changed With the Sampling Interval

Six-Month Frozen Pad Stability Did Not Mean Six-Month Breath Detection

The six-month experiment tested cannabinoids stored on fortified collection pads after preparation for laboratory evaluation. It did not test how long a living person continues to exhale cannabinoids after use. The study followed a collected material Fortified pads at low and high control concentrations were stored at −20°C. After six…

Spread the love

Read Article: Six-Month Frozen Pad Stability Did Not Mean Six-Month Breath Detection

Speak with an expert today!

Contact the offices of Okorie Okorocha for professional and reliable advice which you can trust.

Call (424) 363-3347 Contact Us