A Sensitive Plasma THC Assay Still Needed Additional Forensic Validation

A low detection limit does not establish that an assay has answered every forensic identification question. A 2017 plasma cannabinoid method illustrates this clearly: the authors reported strong quantitative performance while expressly identifying validation steps needed for a different use.

What the method was designed to do

Andrenyak and colleagues developed gas chromatography–tandem mass spectrometry for THC, 11-hydroxy-THC, THCCOOH and CBD in human plasma. Their intended application was controlled research in which the administered cannabinoids were known.

The lower limits of quantification were 0.1 ng/mL for THC and 11-hydroxy-THC, 0.25 ng/mL for CBD, and 0.5 ng/mL for THCCOOH. These are analytical performance values in plasma. They are not whole-blood thresholds and do not define impairment.

Quantification and identification need different checks

The investigators evaluated blank plasma from six sources and reported accuracy, precision and recovery experiments. They also collected a second set of mass-spectrometric transitions that could support ion-ratio monitoring.

However, recording another transition is not the same as validating an identification rule based on the ratio between transitions. The paper explained that its procedures met many, but not all, forensic validation requirements. Ion-ratio validation and a fuller assessment of interference from other drugs remained relevant before extending the assay to forensic unknowns.

How to use the limitation fairly

The correct reading is not that the published concentrations are automatically unreliable. The method was developed for a stated research purpose and demonstrated performance relevant to that purpose. The limitation concerns transferring that evidence to cases with uncertain exposure histories and potentially interfering substances.

A case review should therefore examine the laboratory’s own current validation, its identification criteria, the specimen matrix and the actual chromatographic evidence. Later work by a laboratory may address issues that the original publication did not.

Even complete identification and accurate quantification leave a further question: what does the concentration establish about the person’s condition? Our explanation of THC blood levels and impairment separates the chemical measurement from that functional conclusion.

Primary research

David M. Andrenyak and colleagues. Journal of Analytical Toxicology (2017). Determination of ∆-9-Tetrahydrocannabinol (THC), 11-hydroxy-THC, 11-nor-9-carboxy-THC and Cannabidiol in Human Plasma using Gas Chromatography–Tandem Mass Spectrometry.

Spread the love


The National Black Lawyers

top 40 lawyers

civil trial law

Lawyers of Distinction

Loading...

Recent Blog Articles

Other Medications Changed THC/CBD Spray Exposure in Different Directions

The effect of another medication on THC/CBD exposure depends on the drug and the pathway involved. A human interaction study found lower cannabinoid exposure with rifampicin, higher exposure with ketoconazole, and no significant pharmacokinetic change with omeprazole under the tested conditions. Three separate crossover comparisons Stott and colleagues studied 36…

Spread the love

Read Article: Other Medications Changed THC/CBD Spray Exposure in Different Directions

An Intermediate CBD Stress Response Was Not a Significant Placebo Difference

A CBD group’s average response can fall between placebo patients and healthy controls without demonstrating a significant CBD-versus-placebo benefit. A study of social stress in people at clinical high risk of psychosis illustrates this statistical distinction. Stress hormones and subjective symptoms were separate endpoints Participants at clinical high risk received…

Spread the love

Read Article: An Intermediate CBD Stress Response Was Not a Significant Placebo Difference

Two-Dimensional Cannabis GC–MS Testing Needs Correctly Timed Transfer Windows

Adding a second chromatography column can improve cannabinoid separation while introducing a new failure point: the analyte must reach the second column during the correct transfer window. A plasma assay study showed why those windows require verification. More separation brought more timing requirements Karschner and colleagues developed two-dimensional gas chromatography–mass…

Spread the love

Read Article: Two-Dimensional Cannabis GC–MS Testing Needs Correctly Timed Transfer Windows

Speak with an expert today!

Contact the offices of Okorie Okorocha for professional and reliable advice which you can trust.

Call (424) 363-3347 Contact Us