A Half-LLOQ Check Was Not the Same as Independently Determining Every Detection Limit

The neonatal cannabinoid study explicitly stated that each analyte’s detection limit was not specifically determined. It checked performance at half the lower quantification limit instead.

The validation described what was actually tested

The researchers verified relevant acceptance criteria and a signal-to-noise ratio of at least three at concentrations equivalent to half the LLOQ. Quantification at the LLOQ required more demanding criteria, including accuracy, imprecision and chromatographic identification requirements.

A tested concentration does not locate the exact boundary

Showing acceptable behavior at one low concentration is useful. It does not establish the precise lowest concentration at which detection would remain reliable under every condition. The half-LLOQ value should therefore not silently become a separately estimated universal LOD.

Reporting categories need their definitions

Detection, numerical quantification and clinical interpretation are distinct. A low analytical signal does not automatically support a precise concentration, exposure date or prognosis. The paper’s paired meconium and cord methods also had different measurement ranges. Retaining the actual validation description allows a fair assessment without exaggerating sensitivity or dismissing useful evidence. The supported claim concerns demonstrated low-level performance under the specified protocol, rather than an independently established minimum for every target and specimen.

Related analysis: the study’s broader findings and limitations.

Research cited

Jensen, Wu and McMillin (2019). Detection of in utero Exposure to Cannabis in Paired Umbilical Cord Tissue and Meconium by Liquid Chromatography-Tandem Mass Spectrometry.

Spread the love


The National Black Lawyers

top 40 lawyers

civil trial law

Lawyers of Distinction

Loading...

Recent Blog Articles

Carryover Testing Challenged the Analytical Sequence, Not the Newborn’s Exposure History

The neonatal assay’s carryover assessment examined whether a preceding high-concentration sample could affect a later measurement. It addressed an instrument-process question, not a biological exposure pathway. The protocol deliberately stressed the sequence The investigators used samples at twice the upper quantification limit followed by low-range samples. They considered contamination associated…

Spread the love

Read Article: Carryover Testing Challenged the Analytical Sequence, Not the Newborn’s Exposure History

The Paired Cord-and-Meconium Study Was Not a Survey of All Births

The forty-six specimen pairs came from archived cases in which both tests had been clinically ordered. Their cannabinoid-positive proportions were not population prevalence estimates. Selection followed the availability of paired orders Investigators sought residual meconium and cord tissue from the same newborn where testing for both matrices had been requested….

Spread the love

Read Article: The Paired Cord-and-Meconium Study Was Not a Survey of All Births

A Half-LLOQ Check Was Not the Same as Independently Determining Every Detection Limit

The neonatal cannabinoid study explicitly stated that each analyte’s detection limit was not specifically determined. It checked performance at half the lower quantification limit instead. The validation described what was actually tested The researchers verified relevant acceptance criteria and a signal-to-noise ratio of at least three at concentrations equivalent to…

Spread the love

Read Article: A Half-LLOQ Check Was Not the Same as Independently Determining Every Detection Limit

Speak with an expert today!

Contact the offices of Okorie Okorocha for professional and reliable advice which you can trust.

Call (424) 363-3347 Contact Us