Reading the Reported Dilution Results in a Cannabinoid Saliva Assay

The saliva assay tested dilution of high-concentration samples, but its reported results should not be reduced to a blanket statement that every dilution met the stated acceptance criterion.

High samples were diluted into a defined matrix

The method-development study evaluated 1:2, 1:50 and 1:200 dilutions using blank oral-fluid–buffer mixtures. This challenged whether samples above the ordinary calibration range could be prepared for measurement through a specified procedure.

The investigators stated a dilution-integrity criterion of results within 20% of the target. In the results, reported accuracy ranged from 85.1% to 97.8% of target for Quantisal preparations and from 77.1% to 105.8% for Oral-Eze preparations, with three replicates for each dilution factor.

The lower reported value deserves qualification

A value of 77.1% of target is below the 80% lower boundary implied by that criterion. The reported overall range does not identify, in that sentence, which individual analyte and dilution produced the lowest value. Assigning it to a particular cannabinoid without the underlying results would add unsupported precision.

The narrow conclusion is that the study investigated large dilutions and reported a range that includes a result outside its stated criterion. That is more informative than treating the entire exercise as either uniformly successful or wholly unusable.

Applying this to a high cannabis-test result

A high specimen concentration calls for the actual dilution factor, matrix, corrected calculation and laboratory acceptance decision. The validation concerns quantification of cannabinoids in oral fluid; it supplies no impairment threshold and no universal authorization to dilute any specimen by any factor.

Related analysis: collection buffers and cannabinoid quantification.

Research cited

Concheiro, Lee, Lendoiro and Huestis (2013), Simultaneous Quantification of THC, THCCOOH, CBD and CBN in Oral Fluid by MicroFlow LC–High Resolution Mass Spectrometry. DOI: 10.1016/j.chroma.2013.04.071.

Spread the love


The National Black Lawyers

top 40 lawyers

civil trial law

Lawyers of Distinction

Loading...

Recent Blog Articles

CBN Served as the Internal Standard in a Postmortem THC Assay

The postmortem GC–MS method used cannabinol as an internal standard. CBN was a reference signal in this procedure, rather than a second reported measure of parent THC or a stable-isotope-labeled copy of it. Identify the reference compound Paknahad and colleagues described adding CBN internal standard during preparation of urine for…

Spread the love

Read Article: CBN Served as the Internal Standard in a Postmortem THC Assay

Hydrolysis Order Remained Untested in the Urinary Cannabinoid Protocol

The researchers used combined enzymatic and basic hydrolysis but did not evaluate whether reversing their order would change the result. A successful preparation procedure does not establish that alternative sequences are equivalent. Preparation defined the measured material The urinary CBD study analyzed several parent cannabinoids and metabolites by LC–MS/MS. Conjugated…

Spread the love

Read Article: Hydrolysis Order Remained Untested in the Urinary Cannabinoid Protocol

The CBD Excreted-Dose Calculation Used Volume and Molecular Weight

The study’s percentage of dose excreted was a calculated quantity using urine volume, analyte concentration and metabolite molecular-weight adjustment. It was not simply a urine concentration expressed as a percentage. The calculation had several stages For each collection, the investigators combined measured urine volume and concentration to estimate analyte amount….

Spread the love

Read Article: The CBD Excreted-Dose Calculation Used Volume and Molecular Weight

Speak with an expert today!

Contact the offices of Okorie Okorocha for professional and reliable advice which you can trust.

Call (424) 363-3347 Contact Us